SPECIFICATIONS |
Specifications apply after 30 minute warm-up, ambient 5°C to 40°C |
FREQUENCY |
Frequency Range: |
10MHz to 1000MHz |
Setting Resolution: |
1kHz by direct keyboard entry, or in user-set increments of 1kHz to 999·999MHz by rotary control or increment–decrement keys. |
Display Resolution: |
1kHz |
Accuracy: |
± 2 ppm over temperature range 5°C to 40°C. |
Stability: |
<1ppm/year ageing. |
Phase Noise: |
-110dBc/Hz at 25kHz offset, 500MHz carrier. |
Residual FM (FM Off): |
: |
Equivalent peak deviation for 300Hz to 3·4kHz B/W: |
10Hz at 100MHz carrier, 35Hz at 500MHz carrier, 180Hz at 1000MHz carrier |
OUTPUT LEVEL |
Output Level Range: |
–127dBm to +7dBm (0·1µV to 500mV into 50W). |
Setting Resolution: |
0·1dB (or 0·01µV to 1mV) by direct keyboard entry, or in user-set increments of 0·1dB to 100dB (or 0·01µV to 100mV) by rotary control or increment–decrement keys. |
Accuracy: |
Better than ± 2dBm, except for output levels <-70dBm at 500 -1000 MHz, ± 3dBm. |
Harmonics: |
Typically <-25dBc, maximum –20dBc, any carrier frequency, output level <0dBm. |
Non-Harmonic Spurii: |
<–60dBc at >8kHz offset. |
Carrier Leakage: |
<0.5uV generated into a 50W load by a 2 turn 25mm loop, at 25mm from the generator with output set to <-10dBm into a 50W sealed load. |
Output Impedance: |
50W |
Output Connector: |
TYPE N |
Reverse Protection: |
50V DC |
Output Switch: |
RF OUT on-off switch with LED for ON status. |
FM MODULATION |
Peak Deviation: |
0·5kHz to 100kHz. |
Setting Resolution: |
0·5kHz by direct keyboard entry, rotary control or increment–decrement keys. |
Modulation Frequency: |
Internal 1kHz; External 300Hz to 50kHz |
Deviation Accuracy: |
<±10% of setting ±0.5kHz, excl. residual FM, for 1kHz modulation, internal or 1Vrms external. |
External Modulation |
|
Frequency Response: |
± 1dB from 30Hz to 50kHz relative to 1kHz. |
Distortion: |
<2% total harmonic distortion at 1kHz mod. frequency, 100kHz deviation and 500MHz carrier. |
Input Impedance: |
100kOhm |
Input Connector: |
BNC |
INTERFACES |
Full remote control facilities are available through the RS232 (standard) or optional GPIB interfaces. |
RS232: |
Variable Baud rate, 19200 Baud maximum, 9-pin D-connector. Compatible with Thurlby Thandar ARC (Addressable RS232 Chain). |
IEEE-488: |
Conforming with IEEE488·1 and IEEE488·2. |
GENERAL |
Display: |
20 character x 4 row backlit alphanumeric LCD |
Data Entry: |
Keyboard selection of frequency, amplitude, etc.; value entry by numeric keys or by rotary control. |
Stored Settings: |
Up to 9 complete instrument set-ups may be stored in battery-backed memory. |
Size: |
3U (130mm) height; half-rack (212mm) width; 330mm long. |
Weight: |
4·6 kg. (10 lb) |
Power: |
230V, 115V or 100V nominal 50/60Hz, adjustable internally; operating range ±14% of nominal; 30VA max. Installation Category II. |
Options: |
IEEE-488 interface; 19 inch rack mounting kit. |
ENVIRONMENTAL & SAFETY |
Operating Range: |
+5°C to 40°C, 20-80% RH. |
Storage Range: |
–20°C to + 60°C. |
Environmental: |
Indoor use < 2000m, Pollution Degree 1. |
Safety: |
Complies with EN61010-1. |
EMC: |
Complies with EN61326 |